SX-I310T laser dust particle counter (28.3l / min) is used to measure the size and number of dust particles in the unit area of clean environment. It can directly detect the clean environment of class A and class B clean environment.
SX-I310T laser dust particle counter adopts semiconductor laser light source and large touch screen display, which provides simple operation mode, fast sampling, light and easy to carry, which can effectively save the operator's time. The lightweight, streamlined stainless steel shell and ergonomic handle make it easy for the user to lift it with one hand and place it in any suitable measuring place.
In addition, the data can be downloaded and stored by computer software.
Products are widely used in pharmaceutical, optical, chemical, food, cosmetics, electronic health, biological products, aerospace and other industries of clean environment detection.
Product performance:
1. 28.3l/min (1cfm)
2. Both AC and DC
3. Touch screen control operation
4. Six channel particle size display on the same screen to display the particle concentration (particles / m3) 5. Automatic UCL calculation.
Product features:
1. Using the original Japanese laser diode, the service life is longer and the performance is more stable;
2. 7 Inch Touch screen control, human-computer interaction interface beautiful, simple and easy to operate;
3. All levels of user function management, management more orderly and effective;
4. Large flow can save sampling time.
Technical parameters:
Sample size | 28.3L/min ±5% (1 CFM) |
Measuring particle size range(um) | 0.3,0.5,1,3,5,10μm |
Particle size distribution error | 0.5μm、Particle size distribution error ± 30% |
When the particle counter is in normal working state, the particle concentration indication error of 0.5 μ m particle size is less than ± 10%. | |
clean-down capability | ≤10min |
Relative deviation of measurement repetition | Under the same measurement conditions, the repeatability of continuous measurement of particle concentration is ± 10% |
Display screen | Color 7-inch touch screen |
Laser light source | laser diode |
Inlet air pump | Stable continuous operation, energy saving and environmental protection |
Warm and humid | Matching |
USB | |
Battery working time | 4th |
When the battery is charged | 8th |
Battery form | Lithium batteries are rechargeable, AC 220V 50Hz DC14.8V 36AH |
texture of material | SUS304 stainless steel |
Test method | Single / continuous / delayed |
Sampling time | User set (1 ~ 14400) seconds |
Sampling delay | User set (1 ~ 255) seconds |
Sampling points (position) | Setting range: 2 ~ 9 (confidence detection) 2 ~ 50 (arithmetic average) |
Data storage | 1 ~ 5000 groups, which can be inquired on the screen |
Location label | 0~999 will be marked on the printed data |
Printing mode | Built in thermal printer |
Determination standard of cleanliness grade | ISO14644-1、GMP dynamic and GMP static |
User rights | User parameter setting permissions, query permissions, data export permissions |
Machine size | 291×265×280(mm3) |
weight | About 9.15kg |
Operating environment | Temperature 0 ~ 40 ℃, humidity 10 ~ 70% RH |
Storage environment | Temperature - 30 ~ 45 ℃, relative humidity 0 ~ 90% pH |
enclosure | Sampling port rubber sleeve, isodynamic sampling head, self-cleaning filter, sampling tripod, printing paper, portable protection box, charging adapter, export data CD, USB data cable, anti-static black hose (inner diameter of 6mm), operation manual, original factory calibration report and certificate. |
Audit trail function | It meets FDA 21cfr Part 11 requirements, involving access security, audit trail, data storage and retrieval, and electronic signature. |